The IVF On Demand OnLine Filter Provides protection for 6 months.


Filter Specifications and Certifications

  • Purifies CO2 and N2 gasses
  • VOC Capacity: 1400 (mg) tested with Isopropyl Alcohol, ethanol, methanol, acetone, Toluene, Xylenes, 2-butanone, styrene acrolein, benzene, ethyl acetate, 111-trichloroethane, n-Hexane, Methylene chloride, propionaldehyde, butyraldehyde, hexanaldehyde
  • 2 filters: Activated carbon and HEPA (1 Micron)
  • Fitting connections: Male/Female :1/4” Flow cut-off valve quick coupling with metal latch (2QV/2QV) 
  • Sterilized
  • Regulatory: Certified by CE class 1, meets 93/42/EEC Medical Device Standards
  • Constructed with polypropylene resins and filtration media with 21 CFR Part 177 and USP Class VI biological test for plastic.
  • Maximum operating conditions: 60 o C (140oF), 5.5 bar (80 psi)

Main application and target market

  • Purifies CO2 and N2 gasses
  • Market: Medical
Materials of Construction:
(Standard Construction)
Media: Fibrous Activated Carbon and HEPA media
Media Supports: Polypropylene
Cage, Core, End Caps: Polypropylene, Silicone, Thermal End Cap
Fitting Connections: Male/Female : 1/4” Flow cut-off valve quick coupling with metal latch (2QV/2QV)
Nominal Dimension: Length: 1.5” (38 mm), 2.5” (63 mm), 5” (127 mm), 7.5” (190 mm), and 10” (254 mm)
Diameter: 2.8” (73 mm)
Effective Purification Volume:
(Single Layer/PP Construction)

1.5”: 0.65 grams 
2.5”: 1.3 grams
5.0”: 2.7 grams
7.5”: 4.0 grams
10”: 5.2 grams

Operating Conditions:

Maximum Operating Pressure: Liquid: 5.5 bar (80 psi) at 72 °F/22 °C

Gas: 4.1 bar (60 psi) at 72 °F/22 °C

Maximum Forward Differential Pressure: 4.1 bar (60 psi) at 72 °F/22 °C

Maximum Reverse Differential Pressure: 2.1 bar (30 psi) at 72 °F/22 °C

Minimum Burst Pressure: 8.3 bar (120 psi) at 72 °F/22 °C

Maximum Operating Temperature: 140°F/60°C

 

 